cylinder 6154 (e-beam welded stage3), motor 6418; stage1 displacer is marked 2083;
displacer tip is machined to cone with final diam of 0.471 to avoid rubbing inside stage3, which is slightly tilted
DATE
MAINTENANCE
24 Jun 2008
dewar removed from C9
30 Jun 2008
opened refrig, cleaned and vacuum baked displacers
noted that Envex sleeve on stage3 displacer is badly cracked
SS section of displacer measures 0.496 to 0.4955 dia
replaced carbon bushings and Kap-Seal; also installed new dowel pins in valve clothespins
new stage3 seal, reuse stage1 and stage2 seals
install prototype welded electrical feedthrough because pin 1 is pushed back on old feedthrough
Dusty repolished the radiation shields, which were badly tarnished
note that slots in base of radiation shield are uncomfortably narrow for wire bundles
removed 3mm mixer and amp, installed HIA-188, WBA13 131D
06 Jul 2008
lab test; temps are 60 / 14 / 3.3 / 3.4 / 3.4 K
09 Jul 2008
installed on C12
09 Jul 2010
removed from C12; stage3 had been running at 5 - 5.5 K
20 Jul 2010
opened refrig, cleaned cylinder and displacers (did not seem exceptionally dirty)
replaced stage3 seal (not worn through), reused stage1 and stage2 seals
stage2 seal thickness ranges from 0.059 to 0.063"
Envex section of stage3 displacer measures 0.4993 at base, 0.475 at tip;
Scotch yoke bushings seem OK